diff options
author | Eugeniy E. Mikhailov <evgmik@gmail.com> | 2024-10-01 17:41:37 -0400 |
---|---|---|
committer | Eugeniy E. Mikhailov <evgmik@gmail.com> | 2024-10-01 17:41:37 -0400 |
commit | 9332101350ab84234731debf56aaba1cd82b13ac (patch) | |
tree | c392864258ef78919b8c7773f62a3fc40ea47714 /qolab | |
parent | 1c3f2d8a1c8153bd787b6747eb012c008ad0ed8f (diff) | |
download | qolab-9332101350ab84234731debf56aaba1cd82b13ac.tar.gz qolab-9332101350ab84234731debf56aaba1cd82b13ac.zip |
draft of Siglent sds800xhd scope
Diffstat (limited to 'qolab')
-rw-r--r-- | qolab/hardware/scope/sds800xhd.py | 69 |
1 files changed, 69 insertions, 0 deletions
diff --git a/qolab/hardware/scope/sds800xhd.py b/qolab/hardware/scope/sds800xhd.py new file mode 100644 index 0000000..6e6f231 --- /dev/null +++ b/qolab/hardware/scope/sds800xhd.py @@ -0,0 +1,69 @@ +""" +Created by Eugeniy E. Mikhailov 2021/11/29 +""" + +from qolab.hardware.scope.sds1104x import SDS1104X +from qolab.hardware.scope._basic import calcSparsingAndNumPoints +from qolab.hardware.basic import BasicInstrument +from qolab.hardware.scpi import response2numStr +from qolab.data.trace import Trace +import numpy as np +import scipy.signal +from pyvisa.constants import InterfaceType + + +class SDS800XHD(SDS1104X): + """Siglent SDS800XHD scope""" + + # SDS1104x has actually 8 divisions but its behave like it has 10, + # the grabbed trace has more points outside what is visible on the screen + vertDivOnScreen = 8 + horizDivOnScreen = 10 + + def __init__(self, resource, *args, **kwds): + super().__init__(resource, *args, **kwds) + self.config["Device model"] = "SDS800XHD" + self.resource.read_termination = "\n" + self.numberOfChannels = 4 + self.maxRequiredPoints = 1000 + # desired number of points per channel, can return twice more + + + @BasicInstrument.tsdb_append + def getTimePerDiv(self): + qstr = "TDIV?" + rstr = self.query(qstr) + # Careful! TDIV? is undocumented for SDS800XHD scope, + # the prescribe command is ":TIMebase:SCALe?". + # But "TDIV?" works identical to SDS2304, i.e. + # Siglent claims that this model should have same commands as SDS1104X + # However response is different. + # For example we got '2.00E-08S' instead 'TDIV 2.00E-08S' + # expected reply to query: '2.00E-08S' + prefix, numberString, unit = response2numStr( + rstr, firstSeparator=None, unit="S" + ) + return float(numberString) + + +if __name__ == "__main__": + import pyvisa + + print("testing") + rm = pyvisa.ResourceManager() + print(rm.list_resources()) + # instr = rm.open_resource("TCPIP::192.168.0.62::INSTR") + instr = rm.open_resource("USB0::62700::4119::SDS08A0X806445::0::INSTR") + scope = SDS800XHD(instr) + print(f"ID: {scope.idn}") + # print(f'Ch1 mean: {scope.mean(1)}') + print(f"Ch1 available points: {scope.getAvailableNumberOfPoints(1)}") + print(f"Sample Rate: {scope.getSampleRate()}") + print(f"Time per Div: {scope.getTimePerDiv()}") + print(f"Ch1 Volts per Div: {scope.getChanVoltsPerDiv(1)}") + print(f"Ch1 Voltage Offset: {scope.getChanVoltageOffset(1)}") + print("------ Header start -------------") + print(str.join("\n", scope.getHeader())) + print("------ Header ends -------------") + # ch1 = scope.getTrace(1) + # traces = scope.getAllTraces() |